AMP-210
AMP-210-EUR
600A AC TRMS
Clamp Multimeter
AMP-220
AMP-220-EUR
600A AC / DC TRMS
Clamp Multimeter
AMP-310
AMP-310-EUR
600A AC TRMS
Clamp Multimeter
AMP-320
AMP-320-EUR
600A AC / DC TRMS
Clamp Multimeter
AMP-210 / AMP-210-EUR
600A AC TRMS Clamp Multimeter
AMP-220 / AMP-220-EUR
600A AC/DC TRMS Clamp Multimeter
AMP-310 / AMP-310-EUR
600A AC TRMS Clamp Multimeter
600A AC/DC TRMS Clamp Multimeter
User Manual
7/2014, 6003318 A
©2014 Amprobe Test Tools.
All rights reserved. Printed in Taiwan
English
AMP-320 / AMP-320-EUR
Limited Warranty and Limitation of Liability
Your Amprobe product will be free from defects in material and workmanship for one year from
the date of purchase unless local laws require otherwise. This warranty does not cover fuses,
disposable batteries or damage from accident, neglect, misuse, alteration, contamination, or
abnormal conditions of operation or handling. Resellers are not authorized to extend any other
warranty on the behalf of Amprobe. To obtain service during the warranty period, return the
product with proof of purchase to an authorized Amprobe Service Center or to an Amprobe
dealer or distributor. See Repair Section for details. THIS WARRANTY IS YOUR ONLY REMEDY.
ALL OTHER WARRANTIES - WHETHER EXPRESS, IMPLIED OR STATUTORY - INCLUDING IMPLIED
WARRANTIES OF FITNESS FOR A PARTICULAR PURPOSE OR MERCHANTABILITY, ARE HEREBY
DISCLAIMED. MANUFACTURER SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL
OR CONSEQUENTIAL DAMAGES OR LOSSES, ARISING FROM ANY CAUSE OR THEORY. Since
some states or countries do not allow the exclusion or limitation of an implied warranty or of
incidental or consequential damages, this limitation of liability may not apply to you.
Repair
All Amprobe returned for warranty or non-warranty repair or for calibration should be
accompanied by the following: your name, company’s name, address, telephone number, and
proof of purchase. Additionally, please include a brief description of the problem or the service
requested and include the test leads with the meter. Non-warranty repair or replacement
charges should be remitted in the form of a check, a money order, credit card with expiration
date, or a purchase order made payable to Amprobe.
In-warranty Repairs and Replacement – All Countries
Please read the warranty statement and check your battery before requesting repair. During
the warranty period, any defective test tool can be returned to your Amprobe distributor for
an exchange for the same or like product. Please check the “Where to Buy” section on www.
Amprobe.com for a list of distributors near you. Additionally, in the United States and Canada,
in-warranty repair and replacement units can also be sent to an Amprobe Service Center (see
address below).
Non-warranty Repairs and Replacement – United States and Canada
Non-warranty repairs in the United States and Canada should be sent to an Amprobe Service
Center. Call Amprobe or inquire at your point of purchase for current repair and replacement rates.
USA:
Canada:
Amprobe Amprobe
Everett, WA 98203
Mississauga, ON L4Z 1X9
Tel: 877-AMPROBE (267-7623)
Tel: 905-890-7600
Non-warranty Repairs and Replacement – Europe
European non-warranty units can be replaced by your Amprobe distributor for a nominal
charge. Please check the “Where to Buy” section on www.Amprobe.eu for a list of
distributors near you.
Amprobe Europe*
Beha-Amprobe
In den Engematten 14
79286 Glottertal, Germany
Tel.: +49 (0) 7684 8009 - 0
www.Amprobe.eu
*(Correspondence only – no repair or replacement available from this address. European customers please
contact your distributor.)
AMP-210 / AMP-210-EUR Series TRMS Clamp Multimeters
CONTENTS
SYMBOL.......................................................................................................... 3
SAFETY INFORMATION.................................................................................. 4
UNPACKING AND INSPECTION...................................................................... 5
MEASUREMENTS............................................................................................ 6
Measuring AC and DC Voltage................................................................. 7
Voltage detection (NCV)........................................................................... 8
Measuring AC and DC Current................................................................. 9
Precise Low-Current Measurement.......................................................... 10
Microamps μA Measurement ................................................................... 11
Measuring Resistance, Continuity and Diode ......................................... 12
Measuring Capacitance and Temperature .............................................. 13
Measuring
&
3-Phase Rotation ................................................... 14
Auto Power Off......................................................................................... 15
SPECIFICATIONS.............................................................................................. 16
ELECTRICAL SPECIFICATIONS ........................................................................ 17
MAINTENANCE AND REPAIR......................................................................... 21
BATTERY REPLACEMENT................................................................................ 21
AMP-210 / AMP-210-EUR Series TRMS Clamp Multimeters
7
8
1
9
AMP-320
2
MAX 600A
600V CAT III
1
3
10
NCV(EF)
OFF
TRMS
1
4
ZERO
PEAK-RMS
1
5
REC
SELECT
HOLD
11
6
12
COM
CAT III
600V
1 Jaw
2 Tactile Barrier
3 Jaw Release
4 Display
5 REC/PEAK-RMS Button
6 SELECT/Backlight Button
7 Antenna for Non-Contact
Voltage Detection
8 Precise Low Current
Measuring Location
9 Indicator of the Jaw Center
for Current Measurement
10 Rotary Switch
11 Data Hold/ZERO button
12 Input Terminals
25
24 23 22
13
21
20
14
15
16
19
17
13
Low battery indicator
20
Data hold
14
Alternative Current (AC)
Direct Current (DC)
AC + DC
21
22
23
Varable Frequency Dive
15
Negative reading
16
Relative zero is active
17
18
24
PEAK-RMS mode
(in-rush current) is active
Continuity buzzer is active
Diode test mode is active
Recording mode is active
MAX: MAX mode is active
MIN: MIN mode is active
Precise low current
measurement mode
AVG: AVG mode is active
25
18 kΩ: KiloOhms
Motor rotation indicator
Phase rotation indicator
Auto-ranging
Hz: Hertz
19 μF: Microfarads
μA: Microamps
A: Amps
V: Volts
Symbols
Application and removal from hazardous live conductors
permitted
W
Caution! Risk of electric shock.
T
The equipment is protected by double insulation or reinforced
insulation.
J
Earth (Ground).
CAT III
B
Caution! Refer to the explanation in this manual.
Overvoltage Category lll is for equipment intended to form
part of a building wiring installation. Such equipment includes
socket outlets, fuse panels, and some mains installation control
equipment.
Alternating Current (AC).
3
F
Direct Current (DC).
N
Battery.
Underwriters Laboratories. [Note: Canadian and US.]
P
Complies with European Directives.
Conforms to relevant Australian standards.
=
Do not dispose this product as unsorted municipal waste.
Contact aqualified recycler.
SAFETY INFORMATION
The Meter complies with:
• UL/IEC/EN 61010-1, CAN/CSA C22.2 No. 61010-1, Pollution Degree 2,
Measurement category III 600 V
• IEC/EN 61010-2-033
• IEC/EN 61010-2-032
• IEC/EN 61010-031 (test leads)
• EMC IEC/EN 61326-1
Measurement Category III (CAT III) is for equipment intended to form part
of a building wiring installation. Such equipment includes socket outlets,
fuse panels, and some mains installation control equipment.
CENELEC Directives
The instruments conform to CENELEC Low-voltage directive 2006/95/EC and
Electromagnetic compatibility directive 2004/108/EC.
� Warning: Read Before Using
To avoid possible electric shock or personal injury:
• Use the Meter only as specified in this manual or the protection
provided by the Meter might be impaired.
• Avoid working alone so assistance can be rendered.
• Never measure AC current while the test leads are inserted into the
input jacks.
• Do not use the Meter in wet or dirty environments.
• Do not use the Meter if it appears damaged. Inspect the Meter before
use. Look for cracks or missing plastic. Pay particular attention to the
insulation around the connectors.
• Inspect the test leads before use. Do not use them if insulation is
damaged or metal is exposed.
• Check the test leads for continuity. Replace damaged test leads before
using the Meter.
• Have the Meter serviced only by qualified service personnel.
• Use extreme caution when working around bare conductors or bus
bars. Contact with the conductor could result in electric shock.
• Do not hold the Meter anywhere beyond the tactile barrier.
• When measuring current, center the conductor in the clamp.
4
• Do not apply more than the rated voltage, as marked on the Meter,
between the terminals or between any terminal and earth ground.
• Remove test leads from the Meter before opening the Meter case or
battery cover.
• Never operate the Meter with the battery cover removed or the case
open.
• Never remove the battery cover or open the case of the Meter without
first removing the test leads or the jaws from a live conductor.
• Use caution when working with voltages above 30 V ac rms, 42 V ac
peak, or 60 V dc. These voltages pose a shock hazard.
• Do not attempt to measure any voltage that might exceed the
maximum range of the Meter.
• Use the proper terminals, function, and range for your measurements.
• Do not operate the Meter around explosive gas, vapor, or dust.
• When using probes, keep fingers behind the finger guards.
• When making electrical connections, connect the common test lead
before connecting the live test lead; when disconnecting, disconnect
the live test lead before disconnecting the common test lead.
• Disconnect circuit power and discharge all high-voltage capacitors
before testing resistance, continuity, or diodes.
• Use only 1.5V AAA batteries, properly installed in the Meter case, to
power the Meter.
• To avoid false readings that can lead to electrical shock and injury,
replace the battery as soon as the low battery indicator (N) appears.
Check Meter operation on a known source before and after use.
• When servicing, use only specified replacement parts.
• Adhere to local and national safety codes. Individual protective
equipment must be used to prevent shock and arc blast injury where
hazardous live conductors are exposed.
• Do not use the Meter if the wear indicator in the jaw opening is not
visible.
• Only use the test lead provided with the Meter or UL Listed Probe
Assembly rated CAT III 600V or better.
Unpacking and Inspection
Your shipping carton should include:
1 Clamp meter
1 Test leads
1 Alligator clip set (AMP-310 and AMP-320 only)
1 Banana plug K-type thermocouple (AMP-310 and AMP-320 only)
1 1.5 V AAA batteries (installed)
1 User manual
1 Carrying case
If any of these items are damaged or missing, return the complete package
to the place of purchase for an exchange.
5
AMP-210
MAX 600A
600V CAT III
NCV
(EF)
AMP-310
AMP-220
MAX 600A
600V CAT III
MAX 600A
600V CAT III
AMP-320
MAX 600A
600V CAT III
NCV
(EF)
NCV(EF)
OFF
TRMS
OFF
TRMS
NCV(EF)
OFF
TRMS
TRMS
OFF
PEAK-RMS
REC
SELECT
HOLD
REC
SELECT
ZERO
PEAK-RMS
HOLD
REC
SELECT
ZERO
PEAK-RMS
HOLD
REC
SELECT
HOLD
COM
COM
COM
COM
CAT III
600V
CAT III
600V
CAT III
600V
CAT III
600V
AMP-210
AMP-210-EUR
AMP-310
AMP-310-EUR
AMP-220
AMP-220-EUR
AMP-320
AMP-320-EUR
MEASUREMENTS
� Warning
To avoid possible electric shock or personal injury:
• When measuring current, center the conductor in the clamp.
• When making current measurements, disconnect the test leads from
the Meter.
• Keep fingers behind Tactile Barrier.
• Use the proper function and range for measurements.
• Disconnect circuit power and discharge all high-voltage capacitors
before testing resistance and diode.
• When using probes, keep fingers behind the finger guards.
• Connecting test leads:
- Connect the common (COM) test lead to the circuit before connecting
the live lead;
- After measurement, remove live lead before removing the common
(COM) test lead from the circuit.
Button
Description
SELECT /
Press SELECT button to select the alternative
measurement function on the rotary switch.
Press SELECT button > one second to turn ON or to
turn OFF LCD backlight. LCD backlight automatically
turns off after approximately 32 seconds.
6
HOLD /
ZERO
Press HOLD to freeze the display reading (
is
displayed) and releases the reading when pressed a
second time.
� Warning
To avoid possible electric shock or personal injury,
when Display HOLD is activated, be aware that the
display will not change when you apply a different
voltage.
Press ZERO (HOLD) > one second to clear last
reading from the display ( is displayed) and establish
a baseline for applicable selected functions.
DC-Zero mode for dc and ac+dc currents. Press ZERO
(HOLD) > one second to activate DC-Zero (display
shows “dc_0” for one second). When nulled value is
larger than ± 5 DC A, an acoustic warning (3 short
beeps) will occur.
REC / PEAK-RMS
Press REC button to activate maximum, minimum and
MAX MIN AVG
average reading memory mode (
is displayed). The meter beeps when MAX and MIN
reading is updated during measurement. Press REC
button again to read the MAX, MIN and AVG reading
in sequence. Press REC button > one second to exit
MAX/MIN/AVG reading memory mode.
Press REC / PEAK-RMS button > one second to enter
is displayed) to capture
PEAK-RMS mode (
inrush current or voltage RMS values (80 ms). Press a
second time > one second to exit.
Note: Auto Power Off is automatically disabled under
MAX/MIN/AVG and PEAKRMS modes.
Measuring AC and DC Voltage
To measure ac or dc voltage:
1. Turn the rotary function switch to
or
2. Press SELECT button to choose measurement function: AC V, DC V,
DC+AC V, Hz or EF(Non contact voltage detection). The display reflects
the chosen function mode. For Model AMP-210 and AMP-220, NCV(EF)
function is designed in an independent rotary switch position. See the
Non-Contact Voltage Detection chapter for details.
3. Connect the black test lead to the COM terminal and the red test lead
to the V terminal. Before connecting the probes to the measurement
points, add any clips to the probes that are necessary.
4. Measure the voltage by touching the probes to the desired test points
of the circuit.
5. View the reading on the display.
6. When measuring AC voltage, press SELECT button to view the
frequency reading on the display.
7
Note:
• AC V (and hence DC+AC V and Hz) function is equipped with digital
low-pass filter, and is capable of dealing with VFD (Variable Frequency
Drives) signals. It also improves AC V reading stability in noisy electrical
environments.
Voltage Detection (NCV)
Non-Contact Voltage Detection:
1. For AMP-210 and AMP-220 turn the rotary switch to NCV(EF). “EF” is
displayed. For AMP-310 and AMP-320, turn the rotary function switch
to
or . press SELECT button to toggle to NCV(EF) mode (“EF” is
displayed).
2. The voltage detection antenna
is located along the top-right
end of the stationary clamp jaw for detecting electric field surrounds
energized conductors
3. Detected electric field signal strength is indicated by a series of bargraph segments on the display and beeper. The stronger the electric
field detected, the more bar-graph segments are displayed and more
intense beep sounds.
8
Measuring AC and DC Current
� Warning
To avoid electrical shock and injury:
• Remove Test Leads before making current measurements.
• Do not hold the Meter anywhere beyond the tactile barrier.
• Do not use the Meter to measure currents above the maximum rated
frequency (400Hz). Circulating currents may cause the magnetic circuits
of the Jaws reach hazardous excessive temperatures.
To measure AC or DC current:
1. Turn the rotary function switch to
or
.
2. Press SELECT button to choose measurement function (AC A, DC A,
DC+AC A, or Hz). The display reflects the chosen function mode.
3. Open the clamp by pressing the jaw release and insert the conductor to
be measured into the clamp. Ensure the jaws are firmly closed.
4. Close the clamp and center the conductor using the jaw alignment
marks.
5. View the current reading on the display.
6. When measuring AC or AC + DC current, press SELECT button to view
the frequency reading on the display.
9
� Caution
During current measurement keep the jaws away from other currentcarrying devices such as transformers, motors or energized wires, as they
may negatively influence accuracy of the measurement.
Precise Low-Current Measurement
� Warning
To avoid electrical shock and injury:
• Remove Test Leads before making current measurements.
• Do not hold the Meter anywhere beyond the tactile barrier.
• Do not use the Meter to measure currents above the maximum rated
frequency (400Hz). Circulating currents may cause the magnetic circuits
of the Jaws reach hazardous excessive temperatures.
To measure AC or DC low-current for small conductors:
1. Turn the rotary function switch to
or
.
2. Press SELECT button to choose measurement function (AC A, DC A,
DC+AC A, or Hz). The display reflects the chosen function mode.
3. Open the clamp by pressing the jaw release and insert the conductor to
be measured into the clamp. Ensure the jaws are firmly closed.
4. Align the conductor at the specified jaw tip area for low-current
measurement
5. View the current reading on the display.
6. When measuring AC or AC + DC current, press SELECT button to view
the frequency reading on the display.
10
� Caution
During current measurement keep the jaws away from other currentcarrying devices such as transformers, motors or energized wires, as they
may negatively influence accuracy of the measurement.
Microamps μA Measurement
The μA DC (
) function on the Meter is primarily for HVAC flame sensor
testing. To test a heating system flame sensor:
1. Turn the heating unit off and locate the wire
between the gas-burner controller and the flame
sensor.
2. Disconnect one of the flame sensor wires.
3. Turn the rotary switch on the Meter to
.
4. Connect the black test lead to the COM terminal and
the red test lead to the μA terminal.
5. Using alligator clips, connect a meter in series by
attaching one alligator clip to the disconnected flame
sensor probe and the second one to the disconnected
control-module terminal.
6. Turn heating unit on and check the reading on the
Meter.
7. Refer to the heating unit documentation for what
the correct reading should be.
11
Measuring Resistance, Continuity and Diode
� Warning
• To avoid false readings that can lead to electrical shock and injury, deenergize the circuit before taking the measurement.
• To avoid electrical shock when testing resistance/continuity/diode in a
circuit, make sure the power to the circuit is turned off and all capacitors
are discharged. Use DC voltage function to check the capacitors are
discharged.
1. Connect the black test lead to the COM terminal and the red test lead
to the Ω terminal.
2. Turn the rotary switch to
.
3. Press SELECT button for desired measurement function.
4. Connect the probes across the circuit or component to be tested.
Resistance measurement: The resistance reading appears on the display.
If the circuit is open or resistance exceeds the Meter’s range, the display
reads “OL”.
Continuity measurement: If the circuit is shorted, the Meter beeps and
shows a reading ≤ 10 Ω. (Beeper ON ≤ 10 Ω, OFF > 250 Ω) If the circuit is
open or resistance exceeds the Meter’s range, the display reads OL.
Testing diode: When testing diode, normal forward voltage drop
(forward biased) for a good silicon diode is between 0.400V to 0.900V.
A reading higher than that indicates a leaky diode (defective). A zero
reading indicates a shorted diode (defective). Display reads “OL”
indicates an open diode (defective). Reverse the test leads connections
(reverse biased) across the diode. The display reads “OL” if the diode
is good. Any other readings indicate the diode is resistive or shorted
(defective).
12
Measuring Capacitance and Temperature
� Warning
To avoid electrical shock and injury:
• When testing capacitor in a circuit, make sure the power to the circuit is
turned off and all capacitors are discharged.
• When measuring temperature, DO NOT apply the temperature probe to
any live conductive parts.
Capacitance
1. Turn off circuit power, then disconnect and discharge the capacitor
before measuring capacitance.
2. Connect the black test lead to the COM terminal and the red test lead
terminal
to the
3. Turn the Meter’s rotary switch to capacitance
.
4. Connect the probes across the capacitor to be tested. When measuring,
be sure to note the correct polarity of the capacitor.
Temperature
The Meter measures temperature in either Celsius (°C) or Fahrenheit (°F).
1. Connect the banana plug type-K temperature probe to the Meter’s
input terminal noting correct polarity of the probe.
2. Turn the rotary switch to
.
3. Press SELECT button to select °C or °F. The display reflects the chosen
temperature mode (°C or °F).
4. Position the probe to take the measurement. The reading appears on
the display.
Note: Type-K mini plug temperature probes can also be used with a
plug adaptor with banana pins to type-K socket.
13
Measuring
&
3-Phase Rotation
Measurement is made through the Meter’s terminals L1/L2/L3. Phase
Rotation directions are indicated as symbolic (LCD segments) movements on
the display. Default mode at
. Press SELECT button to toggle between
and
modes.
: Hi-sensitivity mode for checking phase rotation of Motors detects
relatively low signal outputs generated spinning a motor shaft,.
: Normal-sensitivity mode for identifying phase sequence of 3-phase
MAINS supply.
� Caution
Correct rotation detection relies on solid signal connection to all three test
lead terminals simultaneously. Any loose connection will lead to detection
failure and false indication. To verify signal connection and the correct
rotation indication, swap any of two signal connections l to check for
indication of reverse movement.
Hi-sensitivity mode
for Motors:
Connect the test lead L1/L2/L3 to the motor connections by using probes
and/or alligator clips. Be sure the power supply is removed from the motor.
From the perspective of looking down the shaft of the motor, speedspin
it clockwise to generate sufficient signal strength for rotary detection. If
the meter indicates a clockwise movement, the motor leads connected to
L1, L2 and L3 of the meter are L1, L2 and L3 (also known as R, S and T),
respectively. If the meter indicates a counter-clockwise movement, swap any
of two motor connections and retest.
14
Normal mode
for the MAINS circuit:
Connect the test lead L1/L2/L3 to the 3-phase mains circuit by using probes
and/or alligator clips. If the meter indicates a clockwise movement, the
phases connected to L1, L2 and L3 of the meter are L1, L2 and L3 (also
known as R, S and T), respectively. If the meter indicates a counter-clockwise
movement, swap any two connects between the meter and phases. Then
retest. Connecting the above mentioned L1, L2 and L3 of a motor and that
of the MAINS respectively should get a clockwise motor movement.
Using the Beeper feature on
&
mode:
Press and hold the REC button while turning the rotary switch to
to enable the beeper, the display shows “Enbp”. When making rotary
measurement, if the rotary indication segments indicate a clockwise
movement, the beeper sounds a single long beep per segment cycle. If the
rotary indication segments indicate a counter clockwise movement, the
beeper sounds 3 short beeps per segment cycle.
Auto Power Off
The meter turns off if there is no button pushed, rotary function switch
operation for 32 minutes and/or no specified activities below, where
applicable:
1.) Significant measuring readings of above 8.5% of ranges
2.) Non-OL readings for Resistance, Continuity or Diode function
3.) Non-zero readings for Hz function
4.) Significant movement indication as in Phase Rotation functions
The meter will not enter auto power off mode when under normal
measurements. To turn the meter back on, press the SELECT button and
release, or turn the rotary switch off and on to restart the meter.
15
SPECIFICATIONS
Display
3-5/6 digits 6000 counts
Sensing
True RMS
Polarity
Automatic
Update rate
5 per second nominal
Operating
temperature
32 °F to 104 °F (0 °C to 40 °C)
Relative
humidity
Maximum relative humidity 80% for temperature up to
31°C, decreasing linearly to 50% relative humidity at 40 °C
Storage
temperature
-4 °F to 140 °F (-20 °C to 60 °C), < 80% R.H. (with battery
removed)
Pollution
degree
2
Operating
altitude
≤ 2000 m
Temperature
coefficient
nominal 0.15 x (specified accuracy)/ OC @(0OC to 18OC or
28OC to 40OC), or otherwise specified
Transient
Protection
6.0 kV (1.2/50 µs surge)
Overload
protections
Current & Hz functions via jaws: 600 A ac/dc rms at
< 400Hz
Voltage & 3-Phase Rotation functions via terminals:
660 V dc / 920 V ac rms
Other functions via terminals: 600 V ac/dc rms
E.M.C.
Meets EN61326-1:2006
DC A and DC+AC A Functions, in an RF field of 1V/m:
Total Accuracy = Specified Accuracy + 20 digits at around
405MHz
DC µA and Ohm Functions, in an RF field of 1V/m: Total
Accuracy = Specified Accuracy + 25 digits
Other Functions, in an RF field of 3V/m: Total Accuracy =
Specified Accuracy + 20 digits
Agency
approval
P
Power supply
Two 1.5V AAA Size battery
Power
consumption
Typical 13mA for Current functions; 4.3mA for others
Low battery
indication
Approx. 2.85 V for Capacitance & Hz; approx. 2.5 V for
other functions
Auto power
off
Idle for 32 minutes
Auto power
off power
consumption
5µA typical
16
Dimension
(L x W x H)
8.82 x 3.03 x 1.46 in (224 x 77 x 37 mm)
for AMP-220 and AMP-320
8.62 x 3.03 x 1.46 in (219 x 77 x 37 mm)
for AMP-210 and AMP-310
Weight
254 g (0.56 lb) for AMP-220 and AMP-320 with batteries
installed
208 g (0.46 lb) for AMP-210 and AMP-310 with batteries
installed
Jaw opening
& conductor
diameter
1.37 in (35 mm) max for AMP-220 and AMP-320
1.18 in (30 mm) for AMP-210 and AMP-310
ELECTRICAL SPECIFICATIONS
Accuracy is ± (% reading digits + number of digits) or otherwise specified at
23°C ± 5°C.
Maximum Crest Factor < 2.5:1 at full scale & < 5:1 at half scale or otherwise
specified, and with frequency spectrum not exceeding the specified
frequency bandwidth for non-sinusoidal waveforms.
DC Voltage
Range
Accuracy
600.0V
± (1.0 % + 5 LSD)
Input Impedance: 10 MΩ, 100 pF nominal
AC Voltage (with Digital Low-Pass Filter)
Range
Accuracy
600.0V
± (1.0 % + 5 LSD)
Frequency: 50 Hz to 60 Hz
Input Impedance: 10 MΩ, 100 pF nominal
DC+AC Voltage (with Digital Low-Pass Filter)
(AMP-220 and AMP-320 only)
Range
Accuracy
600.0V
± (1.2 % + 7 LSD)
Frequency: DC, 50 Hz to 60 Hz
Input Impedance: 10 MΩ, 100 pF nominal
PEAK-rms (AC V & AC A for AMP-220, AMP-310 and AMP-320 only)
Response: 80 ms to > 90 %
Continuity
Audible Threshold: ON at ≤ 10 Ω ; OFF at > 250 Ω
Response time: 32ms approx.
17
Resistance
Range
Accuracy
600.0 Ω, 6.000 kΩ, 60.00 kΩ
± (1.0 % + 5 LSD)
Open Circuit Voltage: 1.0VDC typical
Capacitance
Range
Accuracy 1)
200.0 µF, 2500 µF
± (2.0 % + 4 LSD)
1)Accuracy with film capacitor or better
Diode
Range
Accuracy
2.000 V
± (1.5 % + 5 LSD)
Test Current: 0.3mA typically
Open Circuit Voltage: < 3.5VDC typically
DC µA (AMP-310 and AMP-320 only)
Range
Accuracy
Burden Voltage
200.0 µA, 2000 µA
± (1.0 % + 5 LSD)
3.5 mV/µA
Temperature (AMP-310 and AMP-320 only)
Range
Accuracy
-40.0 °C to -10.0 °C
±(1% + 1.5 °C)
> -10 °C to 99.9 °C
±(1% + 0.8 °C)
100 °C to 400 °C
±(1% + 1 °C)
-40.0 °F to 14.0 °F
±(1% + 3.0 °F)
> 14.0 °F to 99.9 °F
±(1% + 1.5 °F)
100 °F to 752 °F
±(1% + 2 °F)
K-type thermocouple accuracy tolerances not included
Precise Low Current AC
Range
Accuracy 1) 2) 3) 4)
60.00 A
± (1.5 % + 5 LSD)
Frequency: 50 Hz to 60 Hz
1) Induced error from adjacent current-carrying conductor: < 0.01 A/A
2) Specified with relative zero mode applied to offset the non-zero
residual readings, if any.
3) Add 10 LSD to the specified accuracy @ < 4 A
4) AMP-210 and AMP-310, not specified @ < 0.2 A
18
Precise Low Current DC (AMP-220 and AMP-320 only)
Range
Accuracy 1) 2) 3)
60.00 A
± (2.0 % + 5 LSD)
1) Induced error from adjacent current-carrying conductor: < 0.01 A/A
2) Specified with DC-zero mode applied to offset the non-zero residual
readings, if any
3) Add 10 LSD to the specified accuracy @ < 4 A
Precise Low Current DC+AC (AMP-220 and AMP-320 only)
Range
Accuracy 1) 2) 3)
60.00 A
± (2.0 % + 5 LSD)
Frequency: DC, 50 Hz to 60 Hz
1) Induced error from adjacent current-carrying conductor: < 0.01 A/A
2) Specified with DC-zero mode applied to offset the non-zero residual
readings, if any
3) Add 10 LSD to the specified accuracy @ < 4 A
AC Current
Accuracy 1) 2) 3)
Range
60.00 A
4) 5)
± (1.8 % + 5 LSD) @ 50 Hz to 100 Hz
60.00 A
4) 5)
± (2.0 % + 5 LSD) @ 100 Hz to 400 Hz
, 600.0 A
, 600.0 A
1) Induced error from adjacent current-carrying conductor: < 0.01 A/A
2) AMP-220 and AMP-320: maximum crest factor < 2:1 at full scale & < 4:1
at half scale
3) AMP-210 and AMP-310: specified accuracy is for measurements made at
the jaw center. When the conductor is not positioned at the jaw center,
add 2% to specified accuracy for position errors
4) AMP-220 and AMP-320: add 10 LSD to the specified accuracy @ < 9 A
5) AMP-210 and AMP-310: add 10 LSD to specified accuracy @ < 6 A, and not
specified @ < 0.2 A
DC Current (AMP-220 and AMP-320 only)
Accuracy 1) 2) 3)
Range
3)
60.00 A , 600.0 A
± (2.0 % + 5 LSD)
1) Induced error from adjacent current-carrying conductor: < 0.01 A/A
2) Specified with DC-zero mode applied to offset the non-zero residual
readings, if any
3) Add 10 LSD to the specified accuracy @ < 9 A
19
DC+AC Current (AMP-220 and AMP-320 only)
Accuracy 1) 2)
Range
60.00 A 3) , 600.0 A
± (2.2 % + 7 LSD) @ DC, 50 Hz to 100 Hz
3)
60.00 A , 600.0 A
± (2.7 % + 7 LSD) @ 100 Hz to 400 Hz
1) Induced error from adjacent current-carrying conductor: < 0.01 A/A
2) Specified with DC-zero mode applied to offset the non-zero residual
readings, if any
3) Add 10 LSD to the specified accuracy @ < 9 A
Frequency Hz
Sensitivity 1) (Sine rms)
Range
600 V
50 V
5.00 Hz to 999.9 Hz
60 A (Precise low current)
40 A
50.00 Hz to 400.0 Hz
60 A, 600 A
40 A
50.00 Hz to 400.0 Hz
Function
Accuracy: ± (1.0 % + 5 LSD)
1)
DC-bias, if any, not more than 50% of Sine rms
&
3-Phase Rotation
measurement:
Voltage Range: 65 V to 600 V (sine wave only)
Frequency Range: 35 Hz to 400 Hz
measurement:
Voltage Range: 0.4 V to 600 V (sine wave only)
Frequency Range: 3 Hz to 400 Hz
Voltage Detection (NCV)
Typical Voltage
Bar-graph Indication
20 V (tolerance: 10 V to 36 V)
-
55 V (tolerance: 23 V to 83 V)
--
110 V (tolerance: 59 V to 165 V)
---
220 V (tolerance: 124 V to 330 V)
----
440 V (tolerance: 250 V to 600 V)
-----
Indication: bar-graph segments and audible beep tones proportional to the
field strength
Detection frequency: 50/60 Hz
Detection antenna: inside the top side of the stationary jaw
20
MAINTENANCE AND REPAIR
If the Meter fails to operate, check battery, test leads, etc., and replace as
necessary.
Double check the following:
1. Replace the fuse or battery if the meter does not work.
2. Review the operating instructions for possible mistakes in operating
procedure.
Except for the replacement of the battery, repair of the meter should be
performed only by a Factory Authorized Service Center or by other qualified
instrument service personnel.
The front panel and case can be cleaned with a mild solution of detergent
and water. Apply sparingly with a soft cloth and allow to dry completely
before using. Do not use aromatic hydrocarbons, gasoline or chlorinated
solvents for cleaning.
BATTERY REPLACEMENT
When battery voltage drops below the value required for proper operation,
the battery symbol (
) appears.
� Warning
To avoid shock, injury, or damage to the Meter, disconnect test leads before
opening case.
Replacing BATTERY follow below steps:
1. Disconnect the test lead probe from measuring circuit.
2. Turn the Meter to OFF position.
3. Remove the screws from the battery cover and open the battery cover
4. Remove the batteries and replace with 1.5V AAA Size (IEC R03). Observe
correct polarity when installing the batteries.
5. Put the battery cover back and re-fasten the screw. Battery: 2 x 1.5 V
AAA (IEC R03)
21
Visit www.Amprobe.com for
•
•
•
•
Catalog
Application notes
Product specifications
User manuals
Amprobe®
www.Amprobe.com
info@amprobe.com
Everett, WA 98203
Tel: 877-AMPROBE (267-7623)
Amprobe® Europe
Beha-Amprobe
In den Engematten 14
79286 Glottertal, Germany
Tel.: +49 (0) 7684 8009 - 0
Please
Recycle