Hello!Thank you for purchasing a new handheld oscilloscopes from Jin Han electronic.In order to properly
use this instrument,please read the full text of this manual carefully before using the instrument,
especially about the "safety precautions" part.If you have finished reading this specification , we
recommend that you instructions for proper custody,together with the instrument placed or placed where
you can access at any time,For future inspection during use.
Copyright Information
Hefei Jin Han Electronic Technology Co., Ltd. All rights reserved.
Jin Han-electronic products by China or other countries patent protection,including been made or pending
patent applications.The Company reserves have to change product specifications and prices right.Jin Han
All rights reserved.Licensed software products are all made of Jin Han,protected by national copyright
laws and international treaty provisions.
If the original purchaser from the date of purchase of the product within two years, the products are sold
or transferred to third parties, the warranty period shall be the date of the products from the original buyer
1
from Jin Han or authorized distributors within two years.Probes and other accessories and fuses are not a
guarantee of protection.
If proven applicable warranty period the product is defective, Jin Han discretion to repair the defective
product without charge for parts and labor costs, or (determined by Jin Han) with an equivalent product to
replace the defective product.Jin Han for warranty parts, modules and replacement products may be new or
reconditioned to like new performance.All components, modules and products culvert replacement will
become the property of Jin Han.
Hereinafter referred to "Customer" means, according to a statement of the rights guaranteed by the
provisions of the person or entity.To obtain service under this warranty, Customer must notify the Jin Han
defects during the applicable warranty period, and for the performance of services to make the appropriate
arrangements.Customer shall be responsible for packaging and shipping the defective product to the
service center designated Jin Han, while prepaid freight and provide a copy of proof of purchase of the
original purchaser.For Hong Kong, Macao and Taiwan compatriots and overseas customers, if the product is
returned to the service center, customers should be responsible for paying all shipping charges, duties,
taxes and any other fees.
This warranty does not apply to any defects due to normal wear and tear accident, machine parts, the
use of the product outside the scope of the provisions of, or improper use or improper or inadequate
maintenance caused by failure or damage.Jin Han no obligation to provide the following services in
accordance with the provisions of this warranty:
a)Jin Han-service personnel to not to damage the product installation, repair or maintenance repairs caused;
b)Repair damage caused by improper use or connection to incompatible equipment caused;
2
c)Any damage or malfunction to use the service personnel is not Jin Han caused battery repair;
d) Maintenance has been modified or integrated with other products (if such modification or integration
increases the time or difficulty of servicing the product).
This is guaranteed by Jin Han-made for this product, used to replace any other express or implied.Jin
Han and dealers refused for a particular purpose of merchantability or fitness for any implied warranty.In the
case of breach of this warranty, Jin Han is responsible for the repair or replacement of defective products is
the sole and exclusive remedy available to customers.No matter whether Jin Han and dealers are preadvised of the possibility of any indirect, special, incidental or consequential damages, Jin Han and
distributors of these damage are not responsible.
General safety requirements
The following safety precautions to avoid injury and prevent this product or any other product connected
thereto damage.To avoid possible danger,be sure to use this product in accordance with the provisions.
Only trained personnel to perform maintenance procedures.
1. Avoid fire and personal injury
● Correct plug and pull
When the probe or test leads are connected to a voltage source Do not accessible.
3
● Properly connected probe
Probe ground and ground potential are the same,Do not connect the ground wire to a high voltage. And
during the test, do not touch the exposed contacts and components.
● View all terminal ratings
Inorder to avoid the impact of fires and excessive current, please view all ratings and markings of this
product. Please refer to the product before connecting the product instructions for more information about
ratings.
● Do not open the machine casing operations
If the cover or panel has been removed, do not operate this product.
● Avoid exposed circuitry
Do not touch exposed connections and components of the boot.
● If you suspect a faulty product,do not operate
If you suspect that the product has failed, Please check it by qualified personnel or contact
jinhandz@foxmail.com.
● Maintain adequate ventilation
● Do not operate in wet conditions
● Do not operate in flammable and explosive environment
● Please keep this product surfaces clean and dry
2. Safety Terms and mark
4
This manual terminology.The following terms may appear in this manual:
WARNING: Warning statements indicate might endanger the
safety of life conditions and behavior.
Note: Note the statement pointed out that this product could
result in property damage and other conditions and behavior.
Terms on the Product: These terms may appear on the product
DANGER: Indicates that there is a direct risk of harm exists near the mark.
WARNING: Indicates a potentially dangerous harm nearby markers.
Note: that of the product, and other potentially dangerous property
Symbols on the Product: These symbols may appear on the product
Warning of high pressure Protective earthing Note: please refer to the manual
5
Measuring ground terminal
Foreword
This manual introduces JHJDS30 series with information about the operation of the digital
storage oscilloscope. Includes the following chapters in this manual:
◆User guide: Simple introduces the portable digital storage oscilloscope front panel, the user
interface, function and probe compensation.
◆ Function is introduced and the operation: The oscilloscope, signal generator, multimeter,
recorder, the center of the serial port function and operation is introduced in detail.
◆Application example: Provide some measuring examples, reference for readers.
◆System prompts and troubleshooting:
◆Service and support:
This manual contains the following four hand-held digital storage oscilloscopes; specific model
is shown in the table below:
6
Type no.
JHJDS3012A
JHJDS3012E
JHJDS3022A
JHJDS3022E
Pass way
With wide
Single
channel+
multimeter
+signal generator + recorder +
serial port center
Single channel + signal generator
+ recorder + serial port center
Dual channel + multimeter, signal
generator + recorder + serial
port center
Dual channel + signal generator +
recorder + serial port center
30Mhz
Real-time
sampling
rate
250MSa/s
The depth of
the storage
30Mhz
250MSa/s
16Kpts
30Mhz
250MSa/s
16Kpts×2
30Mhz
250MSa/s
16Kpts×2
16Kpts
JHJDS30 series handheld digital storage oscilloscope provide users with simple and clear
front panel function, for basic operation. Each letter or symbol of the labeled buttons are
marked with appropriate functions, the user can more intuitive, conforms to the use of
traditional instruments used, the user does not have to spend a lot of time to learn and
7
familiar with can skilled use. To speed up adjustment, easy to measure, the user can press the
"AUTO" button; the instrument is suitable for waveform and gear position.
In addition to ease of use, JHJDS30 series handheld digital storage oscilloscope has faster
measurement task need high performance indicators and powerful. Through 250 Msa/s real-time
sampling, can be observed on the JHJDS30 series handheld digital storage oscilloscope signal
faster. Powerful and analyzes waveform makes it easy to capture and analysis. Clear liquid
crystal display and function of mathematical operations, making it easy for users to more
quickly and clearly to observe and analysis the signal problems.
In addition, JHJDS30 series handheld digital storage oscilloscope function signal generator
and digital channel is integration, the user can according to need to bring up the usual signal
waveform, and can be accessed by users digital channel analysis data, greatly satisfy the user
requirements.
JHJDS30 series digital oscillographic table provides unique waveform display interface and a
variety of measurement functions. Can be widely used in engineering survey, out testing,
automotive electronics, frequency converter, research, design, education and training, and
other fields, including virtual circuit measurement and fault.
From the performance characteristics of given below, can understand the series handheld digital
storage oscilloscope how to satisfy your measurement requirements.
8
◆ Set the oscilloscope, signal generator, multimeter functions in one
oscilloscope/oscillographic table
◆ You can observe signal data through serial port center
◆ Built-in recorder function
◆ A serial port center
◆ With digital channels
◆ 4.3 inch 16-bit true color TFT 400*272
◆ Maximum 30 MHZ bandwidth, sampling rate 250 Msa/s, 16 KPTS storage depth
◆ Support USB firmware upgrade
◆ Have the edge trigger function, can automatically detect support(50Hz—60MHz)
◆ A variety of waveform mathematical functions together
◆ Support Chinese and English menu display
◆ Cursor support time, voltage measurement
◆ Automatic, normal, single trigger mode
◆ Support automatic shutdown function and power saving mode, the time can be set up
◆ The cursor function
◆ Screenshot of waveform can be retained and through the USB into the computer
◆ Bring support to facilitate placement
9
isolated
◆ Through the USB power supply
◆ Super-long standby: single battery can work continuously for 4 hours
JHJDS30 series hand held digital storage oscilloscope attachments:
JHJDS3012E/JHJDS3022E Oscilloscope
JHJDS3012A/JHJDS3022A Oscilloscope
◆The directions for use
◆ The directions for use
◆A copy of the certificate of approval
◆A copy of the certificate of approval
◆One(JHJDS3022E has two)1: 1/10: 1 probe ◆One(JHJDS3022A has two)1: 1/10: 1 probe
◆A clipped
◆ A clipped
◆Two lithium batteries
◆A pair of pens and multimeter
◆A lithium battery charger
◆ Two lithium batteries
◆A lithium battery charger
10
Directory
The first chapter
Introductory guide ................................... 1
1.1
1.2
1.3
1.4
1.5
1.6
General inspection ..................................................... 1
The front panel and the user interface ................................. 2
Holding a storage oscilloscope connected input ......................... 6
Functional check ....................................................... 7
The probe compensation ................................................. 7
Preliminary using oscilloscope display signal automatically ............ 9
The second chapter Oscilloscope operation .............................. 10
2.1 End connections ........................................................ 11
2.2 Automatic measurement .................................................. 11
2.3 Display system ......................................................... 11
2.4 The vertical system .................................................... 13
2.5 The level of system .................................................... 18
2.6 The trigger system ..................................................... 19
11
2.7 The storage system ..................................................... 21
2.8 Auxiliary function ..................................................... 23
2.9 The cursor measurement ................................................. 24
2.10Mathematical function .................................................. 25
The third chapter
Signal generator ...................................... 26
3.1 User interface signal generator is introduced .......................... 27
3.2 Signal generator operating instructions ................................ 27
3.3 Custom waveform signal generator ....................................... 29
The fourth chapter
Multimeter function is introduced and the operation
4.1 Multimeter interface is introduced ..................................... 30
4.2 Multimeter instructions ................................................ 32
The fifth chapter
Recorder and center using the serial port ......... 35
5.1 The use of the recorder ................................................ 35
5.2 A serial port center operations ........................................ 37
The sixth Chapter
Application example .................................. 40
6.1 Simple signal measurement .............................................. 40
6.2 The cursor measurement ................................................ 41
6.3 To capture a single signal ............................................. 42
12
6.4 A serial port center print uclinux start information ................... 44
6.5 Generate sine wave signal .............................................. 45
6.6 Multimeter to measure dc voltage ....................................... 46
The seventh chapter
System prompts and troubleshooting ............... 47
7.1 Message system ......................................................... 47
7.2 Fault handling ......................................................... 47
The eighth chapter
Service and support ................................ 49
8.1 Summary of the warranty ................................................ 49
Appendix A technical specification ......................................... 49
Appendix B daily maintenance and cleaning .................................. 54
13
The First Chapter
Introductory guide
JHJDS30 series portable digital storage oscilloscope is small, lightweight portable instruments.
To provide customers with convenient and easy to operate, the front panel can perform basic
tests.
This chapter mainly elaborates how to perform the following tasks:
△General inspection
△The front panel and the user interface
△Holding a storage oscilloscope connected input
△Functional check
△The probe compensation
1.1 General inspection
When you get a new JHJDS30 series with oscilloscope, it is strongly recommended that you follow
the steps below to check the oscilloscope.
Check whether there is damage caused by transportation:
If it is found that serious breakage or foam cushion packaging carton, please keep, until the
14
whole machine and accessories by electrical and mechanical tests.
Check the attachment:
About providing detail of attachment, in front of this manual in the attachment. You can refer
to this check attachment for missing or damaged. If you find attachment missing or damaged,
please contact in charge of the business of dealers.
Check the machine:
If it is found that the appearance of instrument, the operation of the instrument or failed the
performance test, please contact in charge of the business of selling or dealer. If the
instrument damage caused by the transport, please pay attention to keep the packaging. Notify
the transportation department and in charge of the business of selling or dealers. We will
arrange for you repair or replacement.
1.2 The front panel and the user interface
When you get an oscilloscope, you need to be familiar with its front panel. This chapter will
detail oscillographic table layout and how to use the front panel keys. Please read this
chapter carefully before any action.
The front panel and key layout(picture 1-1):
The front panel of buttons in addition can be displayed on the interface menu allows you to
15
directly use some function. Shown on the interface menu provides all of the advanced features.
Function description (Figure 1-1):
1. display
10. HORI:Display the HORI menu
2. F1~F5:Multi-function selection buttons
11. TRIG:Display the TRIG menu
3. PRW: Power on/off button
12. MATH: Display the MATH menu
4. MENU:The menu button
13. MEAS: Display the measurement menu
5. Adjust the trigger level
14. Adjust the vertical voltage notch
6. Set the time base and horizontal displacement waveform 15.Adjust the horizontal position
7. The cursor keys
16. CH1: Display the CH1 menu
8. CH2:According to channel 2 menu
17. AUTO: Automatically set
9. OK:Identify key
18. RUN:Run/stop
16
Figure 1-1
17
Figure 1-2 screens
18
Display interface, as shown in figure 1-2
Oscilloscope interface description:
1.According to the battery
2.According to state trigger
3.Display the current waveform in memory location
4.Show the level of time
5.Displays the wait time trigger signal
6.According to signal amplitude
7.Display, frequency signal cycle
8.According to duty ratio
9.Display trigger position
10.According to channel 1
11.According to channel 1 voltage notch
12.According to the relative horizontal displacement
13.waveform
19
14.The grid
1.3 Holding a storage oscilloscope connected input
Instructions :
1.The oscilloscope probe
2.Oscilloscope input
3.Oscilloscope digital channel input
4.The signal generator output
5.stents
6.Is clipped
7.Multimeter to test the pen
8.Multimeter input pin,
four round pins are respectively 10 a current,
current mA stall, grounding, voltage
Note: figure 1-3 for single channel
oscilloscope connected input, dual-channel
oscilloscope connected input is from left
20
figure 1-3
to right CH1 and CH2 and CH3, OUT (i.e., channel 1, 2, digital channel and signal generator
output)
1.4 Functional check
First of all to the oscilloscope to do a quick examination to determine whether the instrument
work is very normal
Necessary. Please follow the following steps:
1. After confirm the oscilloscope has been installed on the battery, long press on the
oscilloscope panel "PRW" button,
Until when they hear the buzzer rang it let go.
2. Set the switch on the oscilloscope probe to 10 x and will probe and oscilloscope channel 1
The connection. Aim the probe on the connector slot CH1 coaxial cable splice pieces of the
convex
Press key, can be connected, and then spin to the right to tighten the probe.
3. Place the probe of the probe and grounding clip connected to the corresponding connection
end of the signal generator
(suggest enter 1 KHZ, about 3 v peak value of square wave).
21
4. Set the CH1 menu attenuation coefficient of the probe is 10 x (CH1 - > F3 key order
Select 10 x). Press the AUTO button and wait for a moment, this screen will display frequency
1 KHZ, peak value to 3 v square wave signal.
Open 5. Close the CH1 and CH2, repeat steps 2 and 3.
Figure 1-4 probe set
22
1.5 the probe
1. The safety of the probe
Probe body surrounding protection Settings can protect the fingers to prevent electric shock.
Prior to any measurement, connect the probe to the oscilloscope and grounding terminal.
2. The probe compensation (see the probe specification)
In the first connect the probe to any input channel, this adjustment is required, make the
probe to the input channel. Without compensation correction probe will lead to a measurement
error or mistakes. If the adjustment probe compensation, please click the following steps:
(1) the channel attenuation is set to 10 x, will probe options menu to probe on the switch is
set to 10 x, and the oscilloscope probe connected to channel 1. Such as using the probe hook
head, should ensure that contact with the probe and reliable.
(2) the probe end connected to the signal generator output connector, clamp and signal
generator is connected to ground wire connector, according to channel, and then press the
"AUTO" button (automatically).
(3) Check shows the shape of the waveform. As shown in figure 1-5.
23
Owe to compensate
(4)
Compensation OK
Spillover compensation
Figure 1-5
If necessary, adjust the probe, repeated operation when necessary.
1.6 Preliminary using oscilloscope display signal automatically
JHJDS30 series hand held digital storage oscilloscope
has the function of signal automatically set. According
to the input signal, can automatically adjust parameters
such as voltage, time base and trigger mode, so as
to achieve the best state of signals. Automatically
set to work, the requirements of the measured signal
frequency is greater than or equal to 50 Hz.
This part mainly discusses the preliminary use
the oscilloscope's customer to do a demonstration of
automatic display signal but not introduce all the functions
24
Figure 1-6 boot screen
of holographic. The specific steps are as follows:
(1) the oscilloscope after equipped with batteries, long press the oscilloscope in the top
right corner of the red "PRW"
Key, until I heard a buzzer sound can let go, oscilloscope into at this time
Boot screen, as shown in figure 1-6
(2) enter the boot screen, can choose instrument by selecting the button F1 ~ F5
Mode, the default is selected "oscilloscope", selected after about 3 s or so
Automatically entered into the oscilloscope operating interface.
(3) will probe from the oscilloscope channel 1 (CH1) and probe to 10 x gear, pull it
After grounding clip to receive the signal
(4) press the "CH1", then press "F3" will probe set to 10 x
(5) will probe needle comes into contact with the need to test signal source, press the "AUTO"
button,
At this point, the oscilloscope is automatically set vertical, horizontal and trigger system,
and will be
Waveform display the best effect of the signal.
25
The second chapter
Oscilloscope operation
The user from the previous chapter have probably learned how to set the JHJDS30 series
oscilloscope from the menu bar. This chapter will detail the oscilloscope each function
Settings. The content of this chapter is as follows:
△ End connections
△ Delta automatic measurement
△ Delta display system
△ Delta vertical system
△ Delta level system
△ Delta trigger system
△ Delta storage system
△ Delta secondary system
△ Delta cursor measurement
△ Delta mathematical functions
26
2.1 End connections
As shown in figure 2-1: from left to right in the order:
CH1、CH2、CH3、OUT
CH1:Used to display signal input connection
CH2:Digital channel input connection
OUT:The output of the signal generator
end connections
Figure 2-1 connection
Note: the figure 2-1 for single channel oscilloscope connector diagram, double-channel
holographic CH1 and CH2 for signal input port, CH3 for digital channel input port.
2.2 Automatic measurement
JHJDS30 series hand held digital storage oscilloscope/oscillographic table with functions of
automatic setting. According to the input signal, can automatically adjust voltage rank, time
27
base and trigger way to best display form.
“AUTO”the function of the button to automatically set the button.
● If multiple channel signal, has the lowest frequency signal channel as a trigger source.
● Found no signal, it will access a signal channel 1, press the "AUTO" button.
2.3 Display system
Function
menu
Display
type
Display
mode
Display
brightness
Set
Vector
Y-T
X-Y
Backlight
Table 2-1 menu display system function
Instructions
Showed by means of connection between the sampling points
According to relative to the time
vertical voltage
Display (no the single channel oscilloscope)
Can be set from 1 to 5
28
scale
(horizontal)
Color
/
Colour or black and white display
Language
/
In Chinese or English
Display system Settings
1. The backlit Settings: press "MENU" button, find the backlight Settings, press "F5" key set
the backlight, namely the brightness.
2. Language: press "MENU" button, find language setting, press "F2" setting language, can be
set up, both Chinese and English display interface.
3. The color Settings: press "MENU" button, find the color Settings, press "F3" button to set
the color, can be set up two display color and black and white
4. Mode: press "MENU" button, find the mode setting, press "F1" setting mode, can be set Y - T
or x-y mode display
5. Long press "MENU" button, can choose interface, enter the instrument function by F1 ~ F5
select button to choose. As shown in figure 2-2
Figure 2-2 Instrument selection interface function
29
2.4
The vertical system
JHJDS30 series handheld digital storage oscilloscope every channel has an independent operating
menu. When press the CH1 and CH2 bond, the operating menu will be displayed at the bottom of
the screen.
To change the CH1 and CH2 vertical Settings, follow these steps:
1. Press the CH1 and CH2 bond, function menu will be displayed at the bottom of the screen.
2. Press the F1 ~ F5 to choose various states.
Now you can see the below menu:
CH1 menu:
CH2 menu:
Figure 2-3 channel 1, 2 menu item
Table 2-2 in detail the function of each menu item
channel
CH1
channel 1
30
According
CH2
open
channel 2
Open the display waveform
close
1X
10X
100X
Disable display waveform
According to the probe, select one of the attenuation
The probe
coefficient to keep vertical deflection coefficient
correctly.
There are three :1X、10X、100X
coupling
ac
Stop the dc component input signal
dc
By the input signal of ac and dc component
Reverse phase open
Open the inverse function
close
Close the inverse function
1.Set the channel coupling
In applying to the CH1 channel signal, for example, measured signal is a square wave signal
containing ac component.
● CH1, for example, enter a contains the dc component of the signal.
Press the "CH1"→"F4" select Settings to communicate. Only allow ac component, dc component
is filter. Waveform display as shown in figure 2-4.
31
● Press the "CH1"→"F4" option set to dc. Ac and dc component were allowed through. Waveform
display as shown in figure 2-5.
Figure 2-4 coupled dc
Figure 2-5 coupled ac
2.Probe the proportion set
In order to coordinate the attenuation coefficient of the probe, needs to adjust the
corresponding probe in the channel operation menu attenuation ratio. If attenuation coefficient
for X10 probe, oscilloscope probe proportion of input channels should be set to 10 x, to ensure
that show the information of gear and measuring data is correct.
According to the following operation adjustment channels menu probes attenuation coefficient of
32
scale:
Press CH1 and CH2 key, press the F3 choose appropriate Settings, the Settings will be saved to you to change the
Settings again. As shown in figure 2-6
● Press "CH1"→"F3" set the probe for 10 x
3.Waveform reverse set
● Press the "CH1"→"F5" will reverse phase can be set to "on" see waveform inverse, as shown
in figure 2-7
Figure 2-6 set probe 10 x
Figure 2-7 waveform inverse
4.Waveform display Settings
By setting allows any channel of waveform display or not.
for example:To set the waveform does not display the CH1
channel
33
● Press "CH1"→"F2", set the CH1 channel does not display
The waveform. As shown in figure 2-8
5.V/vertical adjustment Settings
Figure 2-8
In regulating the vertical v/cell, v/range is 100 mv/div - 50 v/div (probe 10 x), stepping in
1-2-5 ways, or 10 mv/div - 5 v/div 1 x (probe), 1 v/div - 500 - v/div (100 x) probe. Take the
case of CH1 channel:
● Such as setting the vertical direction of 2.00 V/div, To probe is set to 1 x gear as an
example
●Press "
" to adjust the vertical v/case, at the same time can press"
"to move up, down
the entire waveform. As shown in figure 2-9
●Such as setting the vertical direction of 1.00 V/div, operation steps as above. As shown in
figure 2-10
34
●Set the waveform relatively level move up or down. As shown in figure 2-11
Figure 2-9 set 2 v/vertical
Figure 2-10 sets 1 v/vertical
Figure 2-11 move the waveform
2.5 The level of system
System Settings level scale and level trigger point position, horizontal center of the screen
calibration is waveform time benchmark (reference). Change the horizontal scale can lead to
waveform relative to the screen center expansion or contraction. Horizontal position change the
position of the waveform is relative to the trigger point.
Horizontal position of the change is to display waveform changed position relative to the level
35
of trigger point.
● Time base (TIME/DIV)
Use the buttons below to change the base size, says horizontal on each of the big length of
time
● Press the " " button to change the level of scale, in order to enlarge or reduce waveform.
If you want to stop the waveform acquisition, then press "RUN" button can be realized. As
shown in figure 2-12,2-13
Figure 2-12
Figure 2-13
● Press the "
" button to adjust the channel
36
wave forms of horizontal position.
According to the change of the time
base and its amplitude changes. As shown in figure 2-14
Figure 2-14 level mobile waveform
2.6 The trigger system
Trigger determines the digital storage oscilloscope when to begin to collect data and display
waveform. Once the trigger is set correctly, he can convert unstable display into meaningful
waveform. TRIG control menu key;
● Trigger control
Trigger mode: the oscilloscope triggering method for edge trigger.
● Edge-triggered: when a trigger signal to the edge of a given electricity at ordinary times,
trigger. Edge trigger mode is on the edge of the input signal trigger threshold. When "edgetriggered", namely along the falling edge, the trigger in the rise of the input signal.
Table 2-3 edge-triggered menu function table
37
source
slope
CH1
CH2
rising
falling
automatic
triggered
normal
single
Set the CH1 to source trigger signal
Set the CH2 to source trigger signal
Select trigger signal is rising along the trigger
Choose falling edge trigger signal in the trigger
Set up under the condition of no detected trigger can also
collect waveform
Set up only when the trigger condition collecting waveform
Set when detected a trigger to collect a waveform, then stop
Instructions:
Set the source:
1, press "TRIG" display trigger menu, according to the signal input, press "F1" choose "CH1" or
"CH2".
Set the trigger level
2, press the "CH1" button, then press " " button to adjust the lower left panel channel 1
mark; Click "TRIG" and press "
" button on the right side panel again adjusting trigger flag
arrow, according to the trigger marks relative channel 1 position and the current vertical per
division on behalf of the voltage value to set the trigger level size.
38
Set up the slope,
3, press "F3" button to choose the slope for the "up" or "down".
Set the trigger mode:
4, press "F4" choose "automatic", "normal" or "single". As shown in figure 2-15, 2-16
Figure 2-15 normal trigger
Figure 2-16 single trigger
2.7 The storage system
JHJDS30 series hand held digital storage oscilloscope reference waveform can be stored in the
two groups (Ref - A, the Ref - B), 6 set of screen shots to oscilloscope in internal storage.
39
● First find a screen shot interface press "MENU"→"F4" Screen shot is set to "off"
● Reference waveform storage: press the "MATH" find reference
waveform Settings→ "F1" Settings stored reference
waveform Ref - A, then press "OK" button, which shows
"SAVING" reference waveform is saved
● See the reference waveform: according
to the "MATH"→"F1"→"F3" open the
Ref - A reference waveform, as shown in figure 2-17
Figure 2-17 check the reference waveform
● Intercepting waveforms images: press "MENU" can be set to find the screenshot of the page "F4" the screenshot is set to "on", then press "OK" to save screenshots, need to save a
screenshot, press "F5" set the screenshots serial number. Used in this manual waveform images
are from an oscilloscope screenshots, etc.
40
● Low check capture images:
Turn off the oscilloscope first, and then connect the oscilloscope and computer with USB data
cable is good, and at the same time, press the "OK" button and "PRW" button on the screen to
the bright can let go; At this point, the computers will be prompted to have U disk inserted,
open the U disk can see pictures of interception. After the operation, please click on the
computer to exit the U disk mode, then disconnect the cable, long press oscillograph "PRW"
button to turn it off.
2.8 Auxiliary system
● Auxiliary function by pressing the "MENU" MENU and F1 ~ F5 selection function key set.
"MENU" button pop-up auxiliary system feature set MENU. Table 2-4
41
Table 2-4 auxiliary function table menu
The function menu
Directions
voice
Set the voice "on" or "off"
Automatic shutdown
Save electricity
Factory pattern to
restore
Can be set up "ten minutes", "30 minutes after shutdown or"
Never "(Never) to turn it off
Can be set up for 20 seconds, "40 seconds," 60 seconds
"and" enter power saving mode after the "300 seconds" or
"Never" mode
Restore the factory Settings
Settings: press "MENU" find auxiliary function MENU page, → Press "F2" set the voice "on" or
"off".
Set up automatic shutdown: press "MENU" find auxiliary function MENU page, →Press "F3" set the
automatic shutdown time.
Settings save electricity: press "MENU" find auxiliary function MENU page,→Press the "F4" set
time to save power.
42
Set up the factory pattern recovery: press "MENU" find auxiliary function MENU page, press "F1"
→"F5" the recovery of oscillograph.
● MEAS:Press the "MEAS" screen will display the current measurement menu
2.9 The cursor measurement
In this mode, the cursor X or Y way come in pairs, and can be manually adjust the spacing of
the cursor. Reading is the measurement of voltage or time value. Manual cursor measurement,
follow these steps:
●When using the cursor, needs first to set the source into all you have to measure the channel.
Press the "HORI"→"F2" select source channel (in this setting for CH1)
● Press the "F1" choose "time", "voltage" or "off"
● By clicking on the " ","
", "
"," "button to adjust the cursor position. as shown in
figure 2-18.2-19
43
Figure 2-18 cursor voltage
Figure 2-19 cursor measurement time
2.10 Mathematical function
Mathematical function is to display the function of the CH1 and CH2 channel wave forms as
addition, subtraction
Table 2-5 math menu functions
Function menu
set
According
On or off
instructions
Set the waveform operation on or off
44
Operation
CH1+CH2
CH1-CH2
CH2-CH1
Channel 1 waveform + 2 waveform
Channel 1 waveform - 2 waveform
Channel 2 waveform - 1 waveform
● Press the "MATH" button, then press "F2" set the display "on", press "F3" choosing operation
mode.
Note: only
function.
the
The third chapter
dual-channel
oscilloscope
(JHJDS3022A/JHJDS3022E)
have
this
Signal generator
This chapter mainly introduces the interface and how to use the signal generator signal
45
generator. Signal generator interface at a glance, is also very convenient to operate. Content
summary:
△ The user interface signal generator is introduced
△ Signal generator operating instructions
△ Custom waveform signal generator
Table 3-1 generator parameters
waveform
frequency
amplitude
The offset
Duty ratio
Sine wave
1HZ~20MHZ
100mV~4V
+2V
/
Square wave
1HZ~5MHZ
100mV~4V
+2V
0%~100%
Triangle wave
/
1HZ~1MHZ
100mV~4V
+2V
Oblique wave
/
1HZ~1MHZ
100mV~4V
+2V
Custom waveform
/
1HZ~1MHZ
100mV~4V
+2V
3.1 The user interface signal generator is introduced
1.waveform
7.According to amplitude
46
2.frequency
8.According to duty ratio
3.amplitude
9.According to the offset
4.offset
10.According to the waveform
5.Duty ratio
11.x, y coordinates
6.According to the frequency
Figure 3-1 signal generator interface
3.2 Signal generator operating instructions
● Select signal generator: in the boot after press "F2" select "signal generator; Also can be
in when the oscilloscope in other operating interface, long press "MENU" into the oscilloscope
function selection MENU interface, press "F2" choose "generator".
● After entering the signal generator interface, the following formal operation signal
generator.
① Set the waveform type: according to the "F1" button to choose their own needs of signal
waveform, each time you press the "F1" can switch a waveform and displayed on the screen,
47
choose one of waveform, press "OK" to confirm.
② Set frequency: press "F2" button and then through the "
" button to adjust the
frequency size, after set, press" OK "to confirm.
③ Set range: press "F3" button and then through the "
" button to adjust the amplitude
size, after set, press" OK "to confirm.
④ Set offset: press "F4" key and then through the "
offset size, after set, press" OK "to confirm.
⑤ Set the duty cycle: press "F5" key and then through the "
" button to set the
" button to adjust the
size of duty ratio, after set, press" OK "to confirm.
● After good waveform parameter setting, can be observed through the oscilloscope waveform
signal generator output. Example below set up an offset for the "+ 2.00", by "3 v", 10 KHZ
frequency triangular wave are shown in figure 3-2 below:
48
Figure 3-2 triangle wave
3.3 Custom waveform signal generator
● Text custom waveform parameters, such as table 3-2
Table 3-2 custom waveform parameters format
Text parameter format
Directions
Waveform
version 14020 28007 00001 This three data representation of waveform version
information
information, each must have before custom waveform
Waveform
amplitude Such as: 00300 said waveform amplitude is 3 v, can be manually modify
value
the amplitude values
The wave offset
Such as: 00150 said shift + 1.5 V, its head a character representation
of plus or minus, 0 is positive and 1 negative, behind four characters
as an offset, can be manually modify
Waveform duty ratio
Such as: 00050, said waveform duty ratio 50%, can be manually modify
49
Coefficient
frequency division
Coefficient
frequency division
points
data
of
1
of
2
Formula: (((125000 * 0 xa3d)/(65536 * 2)) * 25)/frequency + 1
Formula: ((frequency * (points + 1) * (coefficient of frequency
division 1) * 65536 * 2) / 25000) / 250000
Formula: ((u32) ((u32) (((250000 * 0 xa3d)/(65536 * 2)) *
25000)/frequency)/frequency division coefficient of 1) - 1
Points out, according to the points to fill in the corresponding values
● A custom waveform instructions
① Create a new text document data calculated according to table 3-2 write a text document
format and save, copy the text documents into the oscilloscope. The oscilloscope custom has two
waveforms, can enter the oscilloscope waveform parameters USB interface to view a text document
BX2 BX1, text documents.
② Into the signal generator interface, press the "
"->" " or " " button to call a
stored custom waveform, see the waveform after long press" MENU "-> interface, press" F1 "into
the oscilloscope waveform can be displayed from the definition.
The fourth chapter
Multimeter function is introduced and the operation
Gradually in this chapter introduces the oscilloscopes multimeter functions, provides a basic
50
operation on menu and use examples. Multimeter use 4 mm safety rubber socket input, 10 a, mA,
COM, V Ω input terminals.
Content summary:
△ Multimeter interface is introduced
△ Multimeter instructions
4.1 Multimeter interface is introduced
Figure 4-1 multimeter interface
Table 4-1 Range multimeter
51
Measuring
type
Dc voltage
Ac voltage
Dc current
Ac current
resistance
capacitance
diode
On and off
Amount of routine
600.0mV
6.000V
60.00V
600.0V
1000V
600.0mV
6.000V
60.00V
600.0V
750V
60.00mA/600.0mA
6.000A/10.00A
60.00mA/600.0mA
6.000A/10.00A
600.0
6.000Ω
60.00kΩ 600.0kΩ 6.000MΩ 60.00MΩ
Ω
40.00 nF
400.0nF
4.000uF
40.00uF
400.0uF
4000uF
0V~2.0V
Below 30 Ω 600.0 Ω gear detection, voice
Note: contact with capacitance of the test on both ends of the 30 s
4.2 Multimeter instructions
52
Table 4-2
The keys
F1
F2
F3
F4
F5
RUN
AUTO
multimeter operation keys function description
instructions
Select "dc voltage" file shortcut, ac voltage, dc voltage switch button
Select the "dc" A file shortcut keys, "dc", "communicate A" switch button
Dc mA "selected" file shortcut keys, "dc mA", "communication mA" switch button
Select the "on-off" file shortcut keys, "on-off", "capacity" to switch button
Selected file shortcut "resistance", "resistance" and "diode" switch button
Multimeter run/maintained (HOLD) button
Press set to manual, range can be set up by this button
Press set to automatic
⒈ Dc and Ac voltage measurement
① Take black pens and the instrument ".com "terminal, red pens and pick up the instrument
Ω" V "terminals.
② Long, oscillographic table "PRW" button to boot and then press "F5" button (or if the
oscilloscope in other interface time holding down the "MENU" button to press "F5") to switch to
the multimeter function interface.
③ Press the "F1" button to select the dc voltage, or press the "F1" once again and switch to
53
the "voltage" file.
④ Connect the test pen to voltage or power supply under test, the display value can be read.
At the same time, the red pens and receive polarity can also be through the LCD display
shows the value of the positive and negative. (when the ac voltage measurement non-polarity
display) the instrument defaults to "automatic" range.
2.Dc and Ac current measurements
① Press "F2" or "F3" will block adjustment to the required range
② The black pens and receiving apparatus ".com "terminal, red pens and pick up in" mA ", the
largest terminal input current of 600 mA.
③ The measurement of pens and string into the circuit under test to read values.
Note: if the current signal under test before the unmeasured is unknown, please press "F2" to
gear to the "dc" or "communication A" file.
3.On and off the test
(1)Take black pens and the instrument ".com "terminal, red pens and pick up the instrument
Ω" V "terminals.
(2) by pressing the "F4" button to select "on-off" file.
(3) when the test is less than 30 Ω resistance or lines, the built-in buzzer will sound.
54
4.Capacitance measurement
(1) take black pens and the instruments of ".com "terminal, red pens and pick up the instrument
Ω" V "terminals.
(2) by pressing the "F4" key to switch to the "capacity".
(3) the measurement pens and connect to stay for about 30 seconds on both ends of the capacitor,
can read the capacitance value.
5.Resistance measurement
(1) Take black pens and the instrument ".com "terminal, red pens and pick up the instrument
Ω" V "terminals.
(2) Press "F5" button to select "resistance".
(3) On both ends of the resistance under test to measure the pens and connection, the display
value can be read, also can be set manually if necessary Range.
6.Diode test
1) Take black pens and the instruments of ".com "terminal, red pens and pick up the instrument
Ω" V "terminals.
(2) Press "F5" key to switch to the "diode" file.
(3) Take pens and test in the diode under test or the ends of the line, a reading. (measured
diode readings for diode conduction voltage drop)
55
Note:
A. The instrument shows a forward and reverse pressure drop, when the diode
reverse connection, instrument display value is negative.
B. Diode and on-off range default are "automatic", not manually range.
C. When testing on and off, to ensure "voices" has been open, otherwise, the
buzzer will not sound.
Settings can be carried out in accordance with the following steps:
Long press "MENU" button, then press "F1" into the interface, and then press the
"MENU" button, find the voice operation MENU, press "F2" will be set to "on" sound.
7.Data remain functional
Press the "RUN" button on the instrument, is showing the data will remain on the screen, screen
displays the word "HOLD" among, even if the input signal to change or eliminate, value will not
change.
Note 1: when using a multimeter to measure, the oscilloscope probe do not grounded!
Note 2: please select correct measuring gear and measured again after range!
The fifth chapter
The recorder and center using the serial port
56
This hand-held oscillograph with signal generator function not only,
also integrated with recorder and serial port center function, greatly
convenient for all aspects of customer demand. Recorder can
automatically record the waveform, a serial port center can be realtime monitoring data from the serial port through digital channels and
display on the screen. This chapter, the main content is as follows:
△ The use of the recorder
△ Delta serial center operations
5.1 The use of the recorder
●
First enter the recorder interface: hold down the "PRW" button, enter the boot interface
press "F4" key to enter the recorder interface (or in other interface such as the oscilloscope
can be long press "MENU" -> "F4"). Figure 5-1
●
Enter the interface, the screen in the middle horizontal line in the state of moving from
57
left to right, connect the probe to the oscilloscope, to record the signal, and then contact to
enter record state, press "F3" in "F1" button to stop recording.
●
After recording the waveform record can press "F5" playback waveform. As shown in figure5-2
Note: the recorder to record waveform frequency range of 20 Hz waveform, records need to
access the oscilloscope interface manually adjust the time base for 100 ms above waveform
display.
58
5-1 The recorder interface
5-2 Record the playback
5.2 A serial port center operations
● First enter the oscilloscope interface: hold down the "PRW" button, enter the boot interface
59
according to the "F1" button to enter the oscilloscope operating interface, and then press
"CH2" -> "F3" will gear is set to 1 x. (this step is very important)
●
To probe the gears is set to 1 x.
● Into the heart of the serial interface: long press "MENU" -> "F3". Menu interface is as
follows:
A serial port center menu page(page1/2)
A serial port center menu page(page2/2)
Table 5-1
Menu item
Instructions
60
Serial port
Serial
port,
serial port
Open or close a serial port
Baud rate
1200~512000
TTL
Into system
RS232、TTL
10 16
/
8
9
NO ODD EVEN
1 0.5 2 1.5
SAVE NOSAVE
/
Set the serial port baud rate, can be set to 1200 ~ 512000 a total of 13
kinds of baud rate
Set the communication level mode
Set the data type
Turn to next page menu
Set the data transmission digits
Set the parity bit
Set the stop bit
Set to save or not save the file
Turn to the previous page menu
Data bits
Check digit
Stop bit
File
● A simple test
Here simple test data from the computer serial port, the operation is as follows:
( 1 ) Open the master computer "serial" serial debugging tools, such as set the baud rate to
9600, 8 bits, check the NO, 1 stop bit to send data base "11 22". As shown in figure 5-3
61
(2) is set to the same as (1) at the center of the serial port Settings, by F1 ~ F5 to select
a menu item, and then by pressing the "
" or"
" , selected set parameters, finally press
the" OK "key.
(3)Open the computer serial port, and set to cycle to send (here to send a set to 1000 ms),
press "F1" key to open the serial port serial center
At this point from the display window can observe the hexadecimal computer serial port to send
data "11 22". As shown in figure 5-4
62
Figure 5-3 master serial port
The sixth chapter
Figure 5-4 center of serial interface
Examples of application
63
6.1 Simple signal measurement
An unknown signal observation circuit, rapid display and measure the signal frequency and
amplitude.
● To quickly display the signal, please click the following steps:
(1)Set probe menu attenuation coefficient to 10 x, and set the switch on the probe to 10 x,
connect the probe to the oscilloscope channel 1 (CH1).
(2)Long press "PRW" button, then press "F1" to enter the oscilloscope (or if the oscilloscope
in other functional interface, press "MENU" -> "F1").
(3)Will probe received corresponding clamp circuit, CH1 probe were connected to the circuit.
(4)Press the "AUTO" button.
Digital storage oscilloscope is automatically set to waveform shows best. Based on this, you
can further adjust the vertical and horizontal gear of waveform, until the waveform display
meet your requirements.
● For automatic measurement signal voltage and time parameter
Digital storage oscilloscope can be automatic measurement for most display signal, to measure
the signal frequency and peak value, please click the following steps:
(1)Press the "AUTO" button, can quickly measure the current waveform.
(2)Page, according to the CH1 button that can be observed frequency and peak values.
64
At this point, the frequency and peak values measured values are displayed on the screen on the
right side position. As shown in figure 6-1
6.2 The cursor measurement
The digital storage oscilloscope can be
automatically measured a variety of
waveform parameters. All of the automatic
measurement parameters can be measured
by the cursor. Use the cursor can be
quickly on the waveform on time and voltage
measurements.
● Measuring the square wave signal peak voltage
Figure 6-1 square wave
CH1 channels, for example, to measure the square-wave signal peak voltage, please click the
following steps:
(1)Press the "HORI" button to enter the main base state of the cursor Settings.
(2)Press "F1" button sets the cursor type to "voltage", press "F2" button to set the source to
65
the CH1.
(3)Through the cursor display window to see the cursor 1 and 2 cursor location (relative to the
middle of zero voltage reference is horizontal) and delta (the cursor 2 - light the
differential pressure after 1).
(4)By "
","
"button to adjust the position of the cursor 2"
", "
" buttons to adjust
the position of the cursor 1; And the position of their change and incremental will update the
display the cursor display window in real time. As shown in figure 6-2, 6-3
Figure 6-2 cursor to measure
Figure 6-3 cursor measure
voltage
6.3 To time
capture a single signal
Press "F1" button sets the cursor type
to "voltage", press "F2" button to set
the source to the CH1.
66
Advantages and characteristics of digital storage oscilloscope is possible to easily capture
pulse, burr and other non periodic signal, if capture a single signal, you first need to signal
has certain prior knowledge, to set the trigger level and trigger along. If the signal
uncertainty may be triggered by automatic or normal way to look at the first, to identify the
trigger level and trigger along.
Steps are as follows:
● As aforesaid set probe and CH1 channel attenuation
coefficient to 10 x.
● To trigger the set:
(1) Click CH1 - press "F3" button, set the coupling to the "dc".
(2) press "TRIG" button, display the edge-triggered menu Settings.
(3) under this menu, press "F1" set edge type for the "slope", press "F2" button to set the
source to the CH1, press "F3" button to set the trigger mode for the "single".
(4) according to the "RUN" button, the left upper corner of the screen shows "Ready",
waiting for the signal to conform to the trigger condition. If a signal to set the trigger
condition, namely, sampling time, displayed on the screen. Using this function can be easy to
catch a chance, for example, the sudden low level: sets the edge trigger to drop along the
trigger, press "RUN" button to start waiting, when there is a low level occurs, the machine
67
will automatically trigger and trigger a before and after a period of time waveform recorded.
Triggered by "HORI" can change the location of the horizontal position, resulting in different
lengths of negative delay trigger, is advantageous for the observation of high level before and
after the waveform. As shown in figure 6-4
Figure 6-5 Serial print uclinux system startup
information
Figure 6-4 To capture a single signal
68
6.4 A serial port center print uclinux start information
This example introduces using oscilloscope serial print center uclinux startup information.
First of all, into the oscilloscope interface set channel 2 (CH2) to X1, then probe received
CH2, clamp to a serial port GND, probe by serial port TXD port. Finally, will enter the center
of a serial port baud rate is set to 115200, the data set to decimal and opens a serial port.
When to start the uclinux system can be observed in the screen of oscillograph serial print
uclinux startup information. As shown in figure 6-5
6.5
Generate sine wave signal
This handheld digital storage oscilloscope with function signal generator, the signal generator
can produce sine wave, square wave, triangle wave, oblique wave four kinds of wave shape. Here
signal generator to generate sine signal (3 v, 1 KHZ) flash is as follows:
① Long press "PRW" button, then press "F2" into the signal generator interface (or if the
oscilloscope in other functional interface, long press "MENU" -> "F2" )
② Enter the operation interface, press "F1" set the waveform type, in this setting for sine
wave.
③ Press "F2" button, and then by pressing the "
" button to adjust the signal
frequency size adjustment and press the" OK "key, in this setting is 1 KHZ.
69
④ Press "F3" button, and then by pressing the "
"button to adjust the signal
amplitude size, adjust good after press" OK "button, the set to 3 v
⑤ Press "F4" button, and then by pressing the "
" button to adjust the signal
offset, adjust good after press" OK "button, in this setting is + 0.00.
After all set, finally shows the waveform is shown in figure 6-6, shown in figure 6-7:
Figure 6-6 generate sine wave
Figure 6-7 sine wave
70
6.6 Multimeter to measure dc voltage
● Using automatic (AUTO) to measure dc voltage
① Long press "PRW" button, then press "F5"
multimeter into the interface(Or if the oscilloscope
in other functional interface, long press "MENU" -> "F5" )
② Press the "F1" button to select "dc voltage".
③ The multimeter pens and the dc voltage under test.
As shown in figure 6-8
The seventh Chapter
7.1 Message system
System prompts and troubleshooting
Figure 6-8 DC voltage
SAVEING:When stored waveform or screenshot, the system will have the prompt show, and in the
middle of the screen has a progress bar.
CLOSING: After startup when long press "PRW", when the battery is too low, or when to
oscilloscope set off time, system will have the hint, and in the middle of the screen has a
progress bar.
USER1 RUNNING:Waveform signal generator can run
USER3 NO FILE:Signal generator no waveform file
71
7.2 Fault handling
1. If you press the "PRW" oscilloscope still black screen, no any display, please follow the
steps below:
(1) open the instrument battery back cover, and check whether there is a battery power supply
or leakage, bilge gas, etc
(2) if you have a multimeter to measure the lithium battery voltage on both ends (normal is not
lower than 3.0 V)
(3) check finished, restart the instrument.
(4) if still can't normal use this product, please contact us.
2, after the acquisition signal, the picture does not appear in the signal waveform, please
follow the steps below:
(1) check whether the probe is normal on the signal cables.
(2) check signal cable is normal on the BNC.
(3) check whether the probe is connected to the normal object under test.
(4) check whether the object under test signal.
(5) to collect the signal again.
3, the measured voltage amplitude value is 10 times greater than the actual value or small 10
72
times:
Check whether the channel attenuation coefficient is consistent with actual use of probes
attenuation ratio.
4, a waveform display, but can not be stabilized:
Check whether triggering menu of trigger source set is consistent with the actual signal input
by the channel.
5, press the "RUN" button without any display:
Check whether the trigger menu trigger mode is in a "normal" or "single", and the trigger level
is beyond waveform range. If so, will trigger level center, or set the trigger mode to "AUTO".
6, waveform display ladder shaped:
This phenomenon is normal. May level when the gear is too low, increase the level of time base
can improve horizontal resolution, improves the display.
The eighth Chapter
Service and support
8.1Warranty profile (JHJDS30xxx series portable digital storage oscilloscope)
73
Guarantee the production and sales of products of the company, two years from the date of
delivery in authorized dealer, there will be no material and process defects. As in the
detailed rules of product warranty to prove defective, we will repair or replacement service.
In addition to the profile or guarantee the use of any of the warranty, we do not make any
express or implied, including but not limited to implied warranties of merchantability and
special applicability. The company for indirect, special, or shall not be liable for the
resulting damage.
Appendix A: technical specifications
Unless otherwise specified, all technical specifications are used to set the attenuation switch
as 10 x of the probe and the series oscilloscope. To verify the oscilloscope is in line with
the technical specifications, the oscilloscope must meet the following conditions:
● Oscilloscope must be within the specified operating temperature run continuously for more
than thirty minutes.
● If the operating temperature range, reach or exceed 5 degrees are calibrated, in addition to
the specifications marked "typical", all specifications are guaranteed.
● Low oscilloscope must belong to the factory calibration period.
Technical specifications
74
Input
Input coupling
Input impedance
AC、DC
1MΩ+2%
Maximum
input
voltage
Probe
attenuation
Probe
attenuation
coefficient
setting
Signal
acquisition
system
Sampling
methods
40 v (X1 probe); 400 v (probe X10) measurable voltage 220 v; (probe X100)
1000 v voltage can be obtained
1X、10X
20pF+5pF
1X、10X、100X
Real-time sampling and random sampling
75
The depth of
the storage
Acquisition
mode
Vertical system
Vertical
sensitivity
Vertical
accuracy
Vertical
resolution
bandwidth
Level of system
Real-time
sampling rate
Horizontal
scanning range
Trigger system
16K
Sampling, peak detection
1 x 10 mv - 5 v (probe), 100-50 mv v (probe 10 x) 1 v to 500 v (100 x) probe
(1-2.5-5 step)
+3%
8bit
30MHz
250 MSa/s
10 ns - 50 ms (1,2.5, step 5), 100-5 s ms Scan (Scan)
76
Model
Type
Automatic
detection
Mathematical
operations
Measurement
system
The
cursor
measurement
Measuring way
Measurement
Equipment
Screen
Battery
Size
Signal
generator
Automatic, normal and single
Rising and falling edge
Support (50 hz to 60 MHZ)
CH1+CH2、CH1-CH2、CH2-CH1
The cursor support time and voltage
Automatic ,manual
Amplitude, frequency, duty cycle, cycle
4.3 -inch 16-bit true color TFT 400 * 272
Section 3000 + mA lithium batteries (single battery for 4 hours)
220*128*37(mm)
77
Frequency
Duty ratio
Amplitude
Offset
Recorder
Time base
length
A serial port
center
Baud rate
Data bits
Check digit
Stop bit
Into the system
Sine wave (0 hz - 20 MHZ), square wave (0 hz - 5 MHZ), triangular
wave/oblique wave (0 hz - 1 MHZ)
0% - 100%
100mV - 4V
+0V - 2V
100ms-5s
To be determined
115200-1200
8,9
NO
1,0.5,2,1.5
10,16
78
Multimeter
DC
AC
Dc current
Ac current
Resistance
Capacitance
Diode
On and off
Maximum
resolution
Amount of routine
600.0mV
6.000V
60.00V
600.0V
1000V
600.0mV
6.000V
60.00V
600.0V
750V
60.00mA/600.0mA
6.000A/10.00A
60.00mA/600.0mA
6.000A/10.00A
600.0Ω 6.000kΩ 60.00kΩ 600.0kΩ 6.000MΩ 60.00MΩ
40.00 nF
400.0nF
4.000uF
40.00uF
400.0uF
4000uF
0V~2.0V
Below 30 Ω 600.0 Ω gear detection, voice
6000 點
79
Appendix B: Daily maintenance and cleaning
Daily maintenance
Do not to store your instrument or placed in a liquid crystal display long under
direct sunlight.
Please do not let the spray, liquid and solvent on the instrument or probe, in
order to avoid damage to the instrument and probe.
Please use the battery to charge batteries in time.
Clean
According to the operation often examine instruments and probes. Please click the
following steps to the outer surface of the cleaning equipment:
1. Use soft cloth to wipe the instrument and probe the outside of the dust. Clean
LCD screen, be careful not to scratch the transparent plastic protective screen.
2. Use a piece of soft cloth to wipe wet but not dripping instruments, please
80
remove the battery before wiping. Can use downy cleaner or clean water scrub. Do
not use any corrosive chemical reagent in order to avoid damage to the instrument
and probe.
Warning: before replace the battery, please make sure the instrument is completely
dry, to avoid electrical short circuit damage caused by moisture instrument.