(877) 634-0982
www.digipwr.com
HD160 Series
160 Watt High-Density AC/DC Power Supply
Product Specification
Universal AC Input 90‐264VAC
2” x 4” Footprint
Single Main Output plus Aux. 12V
Key Product Features
• Medical (2 MOPP) Safety Approved
• ITE Safety Approved
• High Density
• Active PFC
• Low Profile (1.06” height)
• High Efficiency 90% typ.
• Convection rated to 100W
• 12V, 0.5A Aux (Fan) Output
• RoHS Compliant
Description
The HD160 Series of open frame switching power supplies utilizes a highly advanced circuit topology to
deliver 160 Watts in an industry standard package that has a 4.00 x 2.00 inch footprint and 1U height.
The series has been designed meet the requirements of Medical, Telecom and Industrial applications and
operates over the universal AC input range. These supplies have active power factor correction (PFC),
flexible output configurations an auxiliary 12V output that can be used to drive a fan, and compliance to
worldwide safety and EMC standards.
Ratings
Safety and EMC
Input Voltage Range
90 to 264 VAC, 47 to 63 Hz or 170 to 370 VDC
• CSA/UL 60601-1-1 3rd Ed. Safety
Output Power—200 LFM Forced Air
160W (5V model is 100W)
• CSA/UL 60950-1 ITE Safety
Output Power—Free Air
100W
Power Factor
0.98 at 230VAC
Efficiency
90%
Output Ripple
1% pk-pk, 0 to 20MHz
Aux. Fan Output
12V, 0.5A
Size
Industry Standard 2 in. x 4 in. x 1.06 in.
• NEMKO EN60601-1/EN60950-1
• CE Mark (LVD)
• EN50022 (CISPR 22) Conducted
Class A
• EN61000-3-2 Class D Harmonics
• EN61000-3-3 Voltage Fluctuations
• EN61000-4-2, 3, 4, 5, 6, 11 Immunity
Model Selection
Model
Output Power
Standby Voltage
HD 160-105
100 Watts
+5V @ 20A
HD 160-112
160 Watts
12V @ 13.3A
HD 160-118
160 Watts
18V @ 8.88A
HD 160-124
160 Watts
24V @ 6.66A
HD 160-148
160 Watts
48V @ 3.33A
Contact factory for other voltage configurations.
HD160 Series
Electrical Specifications
Input
AC Input Voltage
90-264VAC (47-63Hz) or 170-370 VDC
Input Current
2A Max Continuous
Input Reflected Ripple
FCC 68 part 15 Class B
Power Factor Correction
0.98 at 220VAC (typical)
Input Line Protection
3A 250VAC IEC Type
Hold-up Time
>16msec @ Full Load
Efficiency
90% Typical
Leakage
100/200uA @ 115/230 VAC (max)
Output
Line Regulation
± 0.1% for Vin (min.) to Vin (max.)
Load Regulation
V1 = ±1% / V2 = ±5% Max
Adjustment Range
±5% Minimum
Min. Load Requirement
None
Ripple
± 1% (20MHz)
Transient Response
5% Max Deviation For 50% Load Step
Over-Voltage Protection
115-150% (Latched Shut-down)
Turn-On Delay
1 Sec. Max.
Initial Setting Accuracy
± 1%
Over-Current
110-130% of I-Max (Auto-Recovery)
Aux Fan Output
12 VDC @ 500m A
EMC and Safety Certifications
Electromagnetic Compatibility
Electrostatic Discharge
EN61000-4-2, ±4KV Contact / ±8KV Air Discharge
Radiated Susceptibility
EN61000-4-3, 26-1000MHz, 10V/M, 80% AM
EFT / Bursts
EN61000-4-4, ±2KV
Surges
EN61000-4-5, ±2KV Line-Earth, ±1KV Line-Line
Conducted Immunity
EN61000-4-6, 0.15-800MHz, 10V, 80% AM
Voltage Dips
EN61000-4-11, 95% Dip & 10ms, 30% Dip & 500ms
Voltage Interruptions
EN61000-4-11, 95% Reduction, 5s
Fluctuations & Flicker
EN61000-3-3
Safety & Emissions
Safety Approvals
CSA/UL 22.2 No. 60950-1-M90 & 60601-1-M90, NEMKO EN60950-1 / EN60601-1, CE Mark (LVD)
Conducted Emissions
EN S0022 (CISPR 22) Class A
HD160 Series
Environmental Specifications
Operation Temperature
-20 ~ +50C Derate Minus 2.5%/C from +50C to +70C
Storage Temperature
-20C ~ +85C
Cooling
Forced Air Cooling 200LFM, 100W Max Convection
Humidity
Up to 95% RH Non-condensing
Shock & Vibration
0.75G Peak Half Sine, 6 Axes
MTBF
>300,000 Hours
Mechanical Outline
Digital Power Corporation
48430 Lakeview Blvd., Fremont, CA 94538, USA
www.digipwr.com | (877) 634-0982
T: (877) 634-0982 | F: (510) 657-6634
sales@digipwr.com
A Coolisys Technologies Company
Digital Power Corporation designs and manufactures full custom, value added and standard comprehansive power
solutions for the most demanding applications in the defense, healthcare, telecom, and industrial markets.
HD160 V6_190924
很抱歉,暂时无法提供与“HD160-148”相匹配的价格&库存,您可以联系我们找货
免费人工找货