USER GUIDE TB-9089
WSMONITOR2 Plus Verification Tester
Operation Instructions
Made in the
United States of America
Features and Components
A
B
C
D
E
Figure 1. SCS 770076 WSMONITOR2 Plus Verification
Tester
Description
The SCS 770076 WSMONITOR2 Plus Verification
Tester is used to perform periodic test limit verification
of the SCS 770075 WSMONITOR2 Plus Workstation
Monitor. Verification may be accomplished without
removing the monitor from its workstation. The
WSMONITOR2 Plus Verification Tester is National
Institute of Standards and Technology (NIST) traceable.
Frequency of verification is based on the critical
nature of the ESD susceptible items handled. SCS
recommends annual calibration of workstation monitors
and the WSMONITOR2 Plus Verification Tester.
Packaging
1
1
1
1
1
1
WSMONITOR2 Plus Verification Tester
Alligator Clip
10mm Stacking Snap
Ground Plug Adapter
Ground Extension Cord, 5'
Certificate of Calibration
Figure 2. WSMONITOR2 Plus Verification Tester
features and components
A. Mat Test Lead: Connect to the monitor's mat
terminal to verify its mat test circuit.
B. Ground Lead: Connect to equipment ground to
provide a ground reference for the WSMONITOR2 Plus
Verification Tester.
C. Operator Test Lead: Insert into the monitor's
operator jack to verify its operator test circuit.
D. Mat Test Switch: Toggle the rocker switch to the
left to force the monitor's mat circuit to fail. Toggle the
rocker switch to the right to force the monitor's mat
circuit to pass.
E. Operator Test Switch: Toggle the rocker switch
to the left to force the monitor's operator circuit to
fail. Toggle the rocker switch to the right to force the
monitor's operator circuit to pass.
SCS - 926 JR Industrial Drive, Sanford, NC 27332
East: (919) 718-0000 | West: (909) 627-9634 • Website: StaticControl.com
TB-9089 Page 1 of 3
September 2018
© 2018 DESCO INDUSTRIES INC
Employee Owned
Operation
VERIFYING THE OPERATOR CIRCUIT
1. Connect the WSMONITOR2 Plus Verification
Tester's green ground lead to equipment ground.
This may be done using the included Ground Plug
Adapter or alligator clip. See Figure 3.
3. Set the operator test switch to FAIL. The monitor's
red operator 1 LED should illuminate, and its
audible alarm should sound.
4. Set the operator test switch to PASS. The monitor's
green operator 1 LED should illuminate.
5. Insert the verification tester's black operator test
lead into the operator 2 jack of the WSMONITOR2
Plus Workstation Monitor.
6. Set the operator test switch to FAIL. The monitor's
red operator 2 LED should illuminate, and its
audible alarm should sound.
7. Set the operator test switch to PASS. The monitor's
green operator 2 LED should illuminate.
8. Disconnect the operator test lead from the monitor.
VERIFYING THE MAT CIRCUIT
9. Connect the included stacking snap to the
verification tester's white mat test lead.
Figure 3. Using the Ground Plug Adapter to ground the
WSMONITOR2 Plus Verification Tester
10. Disconnect the monitor's black mat monitor cord
from its worksurface mat and turn it over to expose
its 10mm snap.
11. Connect the verification tester's white mat test lead
to the mat monitor cord's 10mm snap.
2. Insert the verification tester's black operator test
lead into the operator 1 jack of the WSMONITOR2
Plus Workstation Monitor.
Mat Monitor Cord of
WSMONITOR2 Plus
Workstation Monitor
Figure 5. Connecting the WSMONITOR2 Plus
Verification Tester to the mat monitor cord of the
WSMONITOR2 Plus Workstation Monitor
Figure 4. Connecting the WSMONITOR2 Plus
Verification Tester to the operator jack of the
WSMONTIOR2 Plus Workstation Monitor
SCS - 926 JR Industrial Drive, Sanford, NC 27332
East: (919) 718-0000 | West: (909) 627-9634 • Website: StaticControl.com
TB-9089 Page 2 of 3
© 2018 DESCO INDUSTRIES INC
Employee Owned
12. Set the mat test switch to FAIL. The monitor's red
mat LED should illuminate, and its audible alarm
should sound.
13. Set the mat test switch to PASS. The monitor's
green mat LED should illuminate.
Limited Warranty, Warranty Exclusions, Limit
of Liability and RMA Request Instructions
See the SCS Warranty -
StaticControl.com/Limited-Warranty.aspx
Specifications
Operating Temperature 50 to 95°F (10 to 35°C)
Environmental
Requirements
Indoor use only at altitudes
less than 6500 ft. (2 km)
Maximum relative humidity
of 80% up to 85°F (30°C)
decreasing linearly to 50% @
85°F (30°C)
Dimensions
3.8" L x 2.4" W x .9" H
(97 mm x 61 mm x 23 mm)
Weight
0.2 lbs. (0.1 kg)
Country of Origin
United States of America
SCS - 926 JR Industrial Drive, Sanford, NC 27332
East: (919) 718-0000 | West: (909) 627-9634 • Website: StaticControl.com
TB-9089 Page 3 of 3
© 2018 DESCO INDUSTRIES INC
Employee Owned